Practical
Care, cleaning and safe handling
Good borosilicate glassware lasts decades. Most of it does not, and the reasons are almost always avoidable: the wrong cleaning bath, an unsupported clamp, or heat applied too fast.
Practical
Good borosilicate glassware lasts decades. Most of it does not, and the reasons are almost always avoidable: the wrong cleaning bath, an unsupported clamp, or heat applied too fast.
Interactive
Pick a residue and get the method that works on it — and the one that will ruin the glass.
Recommended method
The routine
The test for clean. Water should sheet evenly off the inner surface with no droplets clinging. Beading means an oily film is still there, whatever the glass looks like.
Before you pull vacuum
A vessel under vacuum holds roughly 10 tonnes of force per square metre pressing inward. Sound glass carries that easily. Flawed glass does not, and it fails without warning, inward, sending fragments in every direction.
Hold the piece up to a good light and turn it slowly. You are looking for:
Never apply vacuum to flat-bottomed glassware. Erlenmeyer flasks, beakers and media bottles are not designed for it. Use a round-bottom flask, a proper filter flask, or a vessel explicitly rated for vacuum — and shield it.
Specific items
Wash a new frit with dilute hydrochloric acid and rinse thoroughly before first use, to clear loose glass dust from manufacturing. Always clean by back-flushing from the stem side — rinsing from the top drives residue deeper into the channels, where it sets and permanently reduces flow. Never put a frit in a base bath: hot alkali widens the pores and the disc quietly stops meeting its grade. There is more detail in the porosity guide.
Clean grease off with a solvent appropriate to the grease type before washing, or the detergent will simply smear it around. Do not use abrasives on a ground surface, and keep joints out of alkaline baths, which dull the grind and spoil the fit. Store long-term with the joint assembled and lightly greased, or capped, so nothing settles on the ground zone. See the joints guide for greasing and freeing seized joints.
Calibrated volumes are only valid at the stated temperature, and heat changes them permanently. Keep volumetric flasks, pipettes and burettes out of drying ovens above about 150 °C, and never flame them. Clean with detergent, and let them drain dry rather than forcing them. If the water does not sheet evenly down the inside wall, the calibration cannot be trusted either.
Disassemble PTFE stopcocks for cleaning and store them loosened, not tightened — PTFE creeps under sustained load and a plug left clamped tight will deform. Glass stopcock plugs must stay with their own barrel; they are ground as a matched pair and are not interchangeable between valves. Keep O-rings out of solvents they are not rated for; the glass will be fine, the elastomer will not.
Always keep a Dewar in its protective mesh or housing. The vacuum jacket means a failure is an implosion, not a crack. Inspect for any loss of silvering or clouding between the walls, which indicates the vacuum is going. Cool slowly by adding cryogen in stages rather than filling in one pour.
Store on padded shelves, not stacked, and never rim-down on a hard surface. Keep pieces from touching each other — glass against glass during a knock is how chips start. Long items such as condensers and columns should lie supported along their length or hang vertically, not bridge a gap between two points.
Replacing something
Cracked, chipped or etched glassware is not worth the risk. The Synthware range covers the standard sizes.